Quantcast
Channel: Electro help
Viewing all articles
Browse latest Browse all 2033

CARDIOFAX GEM - ECG-9010K - ECG9020K - TROUBLESHOOTING TABLE & SYSTEM ERROR MESSAGES -Medical Equipment

$
0
0

TROUBLESHOOTING FLOW CHART {Click on charts to Enlarge}
TROUBLESHOOTING TABLE
SYSTEM ERROR MESSAGES 
During power-up and operation the instrument continuously checks itself for system failure. If a failure is detected, system information and error history are printed on the recording paper and all operations are stopped. System information and error history are also displayed (ECG-9020 only) or printed due to transient noise. After printing the system information and error history, the power of the instrument is automatically turned off.
NOTE: If the same system information appears again after restarting the instrument, do not use the instrument until service personnel has corrected the cause of the problem. Sending a copy of the system information to your nearest Nihon Kohden distributor helps us to troubleshoot your problem quickly.
ERROR CODES DISPLAYED
  • “Error 05”also appears when any key on the operation panel is pressed and held down.
  • When “Error 08”appears, the following settings are not reset to the factory initial settings even if the instrument is initialized.
display language - hum filter
hospital name - direct/modem connection
recording resolution setting - cue mark position
elapsed time - local language font
saved ECG data
ERROR HISTORY
Indicates the latest three errors and the date of the latest error, as in the example below.
##### System Information #####
Error number: 15
Date: Apr 20, 1998 10:41 AM
Program: 9020K
Version: 01-00 01-01 01-01

##### Error History #####
No.1 Error number: 15 Apr 20, 1998 10:30 AM
No.2 Error number: 15
No.3 Error number: 15


Viewing all articles
Browse latest Browse all 2033

Trending Articles



<script src="https://jsc.adskeeper.com/r/s/rssing.com.1596347.js" async> </script>